Jump to the content of the page

When to use a normalization with a FISCHER X-Ray instrument?

The normalization causes an adjustment of the measurement task for current settings or for new current substrates. This has to be conducted in case of modifications of primary filters or anode currents or collimators. It is also necessary, if alloy compositions or substrates of the sample have changed.

My FISCHER X-Ray instrument seems to measure implausible values. How can I ensure, that I am measuring correctly?

You should then choose measuring equipment monitoring. You will check the measurement instrument by remeasuring the calibration standards. In case the correct values are not met, adjustments are necessary.

What does a reference measurement signify, with a FISCHER X-Ray instrument?

A reference measurement is a new calibration of the energy axis. This is relevant for the correction of proportional counter instruments, on the influence of temperature.

How to check a calibration on the FISCHER X-Ray instrument?

In order to check the calibration, it is necessary to remeasure the calibration standards within the menu position 'Product', 'Measure CAL. standards'. If you find, there is a deviation, then the instrument needs to be recalibrated.

How large is the measuring spot for X-Ray measurements?

The measuring spot is dependant on the collimator and the measuring distance. Typical values are: 30 Β΅m - 3mm.

Radiation protection for the FISCHER X-Ray instruments?

Most of our instruments are full protection instruments with a design approval according to the German X-Ray Ordinance.

What can be measured with the X-Ray method?

One can measure elements from the atomic number 11 and coating thicknesses from  approx. 0,005 Β΅m to 60 Β΅m. This depends on the ambient medium (Air, He, Vacuum), detector, size of measuring spot and the atomic number, and of course the application. 

What does the Mask 'Data export' signify?

With the export mask one can define, which parameters will be exported and when and where the data will be sent to. The measurement data will then be available as a text file.

Is it possible to stack the X-Ray calibration foils during the calibration?

Yes, it is possible. There is a rough rule for usage: For proportional counter instruments one may use 2-3 foils. For instruments with a PIN/SDD-detector it is possible to use one foil.

How often does one need to re-certify FISCHER X-Ray calibration standards?

This is dependant on usage and the customer can determine it by himself.  A typical value would be every 1-3 years.

Does one need to re-certify the plate for pure elements, for the FISCHER X-Ray instruments?

This not necessary. You don't need to re-certify the plate, because the elements are saturation thick and therefore very stable.

How accurate are the measurement results of the Fischer X-RAY devices?

The measurement accuracy may vary depending on the measuring application: It’s a function of the measurement time, the measurement spot and the uncertainty of the standards with which the device was calibrated.

What’s being measured when the device asks for 'Scatt'?

A scattering spectrum is required here. There’s no need to measure the scattering spectrum; it can be loaded from the menu: General β–Ί Load Spectrum and Evaluate…

What should be done if the instrument asks for β€˜Base material of cal. std. set’ and β€˜ProductΒ΄s base material’ during normalization or calibration?

Here, WinFTM is asking for the substrate material. Please place and measure the uncoated base material from the calibration standard set and the uncoated base material of the objects being measured. Caution: if the wrong parts are placed on the instrument, this can seriously impact the correctness of the results!

Why can't I create a new measurement task?

The super software is not activated.

What’s the difference between a manufacturer's certificate and an ISO 17025 certificate for calibration standards?

Calibration standards with the ISO 17025 certificate are measured according to a strict procedure as defined by the accreditation bodies; they have less measurement uncertainty than calibration standards with a manufacturer's certificate.

The device is printing out all measured values without being asked to.

Most probably, the option File β–Ί Print Single Readings has been activated in the menu. In this case, each individual value is sent to the printer buffer; then, once a page is full, it’s printed automatically. Deactivate Print Single Readings and clear the printer buffer.

Some measurement readings were accidentally deleted. Can they be restored?

If single readings in the block were deleted, a dash will appear in the list of measured values. These measurements can be displayed again in the menu Evaluation β–Ί Undelete Reading. However, if all the measured readings of a block or product were deleted, the data cannot be restored.

Jump to the top of the page