How to find the right XRF instrument for your application: the most important criteria for the perfect instrument configuration
Which XRF instrument best meets the testing requirements in your quality control? We show you how to avoid costly purchasing mistakes.
When purchasing a new measuring instrument for X-ray fluorescence analysis (XRF), many quality managers and inspectors are confronted with an enormous variety of models and instrument options. They are right to ask themselves: Which measuring instrument is suitable for my application? How must my measuring instrument be equipped so that I obtain reliable results for my measurement task?
The most important finding up front: The most expensive XRF instrument is not always automatically the best choice for every application. Measurement performance depends to a large extent on how well the individual components are matched to the specific requirements of the application.
In this article, you will learn which criteria are decisive when selecting an XRF instrument in order to achieve the best measurement performance and reliable results without investing in unnecessary costs.

What is an XRF instrument?
An XRF instrument uses X-ray fluorescence analysis (XRF) to analyze the composition of materials and – depending on its configuration – to determine coating thicknesses. It consists, among other things, of two main components:
- an X-ray tube that generates the primary X-rays
- a detector that captures the fluorescence radiation emitted by the measurement object
Depending on the sample material, sample size, sample geometry, measurement spot size, layer system, layer thickness, number of layers and required tolerances, different X-ray tubes, anode materials, X-ray optics and detector types are used. Different measurement directions and degrees of instrument automation are also possible.
The most important selection criteria for your XRF instrument
To obtain reliable results and achieve optimum measurement performance for your measurement task, the following factors are decisive when selecting the XRF instrument:
- X-ray tube and anode material
- Detector type
- X-ray optics and measurement spot size
- Measurement direction and sample positioning
- Measurement environment
- Sample geometry and sample size
- Required coating thickness range
- Required elemental analysis and detection limits
1. X-ray tube and anode material: The optimum excitation
The X-ray tube generates the primary X-rays with which the measurement object is excited. Depending on the measurement task, an XRF instrument can be equipped with a standard or microfocus X-ray tube.
Standard X-ray tube
Ideal for: Classic coating thickness measurements of, for example, zinc on iron (Zn/Fe), nickel on copper (Ni/Cu), or chromium-nickel-copper coatings (Cr(Ni/Cu), where the measurement spot can generally be 1 mm or larger due to the comparatively large sample geometry. This type of tube is typically used for incoming goods inspection and/or in electroplating operations.
Advantages: Reliable, robust and cost-effective
Microfocus X-ray tube
Ideal for: High-precision measurements with small measurement spots and thin layers.
Advantages: Compared with a standard tube, higher primary intensity at comparable measurement spot sizes, resulting in reduced measurement times and lower variation.
Anode material
The anode material of the X-ray tube influences the energy spectrum of the primary X-rays with which the elements in the sample are excited. Depending on the element range to be examined and the required detection limit, the appropriate excitation conditions must be selected.
Available anode materials include:
- Tungsten
- Rhodium
- Molybdenum
- Chromium
Tungsten (W) is used most frequently because it produces a particularly intense and versatile X-ray spectrum. Special anodes: More specialized anode materials such as rhodium (Rh), molybdenum (Mo) or chromium (Cr) are used for special applications.
2. The right detector: key to resolution and measurement range
The detector captures the fluorescence radiation emitted by the measurement object and largely determines which measurement tasks can be performed with the XRF instrument.
Proportional counter tube
Ideal for: Comparatively simple coating thickness measurements in the range of approximately 0.1 µm to 50 µm, depending on the coating system.
Advantages: Thanks to its large active detector area and slightly curved window, a high proportion of the fluorescence radiation can be captured by the detector. This is particularly advantageous when a measurement distance between approximately 20 mm and 80 mm is required.
In addition, a proportional counter tube is less sensitive to deviations in sample alignment, adjustment of the measurement distance and sample geometry.
Silicon PIN diode (Si-PIN)
Ideal for: More demanding coating thickness measurements as well as material analysis.
Advantages: Offers significantly higher energy resolution than a proportional counter tube. A somewhat longer measurement time may be required for very small measurement spots.
Silicon drift detector (SDD)
Ideal for: Extremely thin layers down to the nanometer range, complex multilayer applications and highly precise material analyses.
Advantages: Offers the highest performance and best energy resolution of the detector types described here. The measurement range described extends from sodium under vacuum conditions to uranium. For elements present only in low concentrations, detection limits in the parts-per-million (ppm) range are possible.
3. X-ray optics and measurement spot size: The perfect focus
The X-ray optics limit the cross-section of the primary beam and define the measurement spot size on the sample. As a general rule: For precise measurements, the measurement area should always be larger than the measurement spot. If the sample geometry permits, the measurement spot should be selected as large as possible for optimum precision. For cylindrical parts, for example, the sample diameter should be approximately four times the measurement spot diameter.
Collimators
- Allow measurement spot sizes from several millimeters down to approximately 100 µm, depending on the measurement distance.
- Please note: The smaller the collimator, the less primary X-ray radiation reaches the sample, which weakens the fluorescence signal.
Polycapillary optics
- Allow tiny measurement spots down to approximately 10 µm.
- They consist of hundreds of microscopically fine glass capillaries that are drawn into a defined shape and then bundled. Polycapillary optics can focus a large proportion of the primary radiation onto a tiny point and are therefore frequently used in the electronics and semiconductor industries to measure delicate structures.
4. Measurement direction and positioning
The geometry of your samples determines the optimum measurement direction:
Measurement from top to bottom
- Allows measurement of samples with heights of up to 80 mm as well as components with recesses.
- A laser pointer supports precise positioning of the sample.
- A programmable XY stage enables rapid movement between different measurement points on multiple samples and also allows semi-automation of your measurements.
Measurement from bottom to top
- Ideal for flat samples such as sheets, coins, jewelry or precious-metal bars.
- Measurement distances of up to 35 mm can be achieved without difficulty.
Flexible measurement direction
- Handheld XRF instruments enable measurements on large, bulky parts such as aircraft carriers, turbine blades or pipelines.
5. The right measurement environment: From air to vacuum
The measurement environment influences which elements can be analyzed and which detection limits can be achieved.
Air
Air is the simplest and most cost-effective measurement environment and is suitable for the vast majority of applications. However, scattering and absorption of the fluorescence radiation can occur in air. For most heavy elements, air absorption can be neglected.
Helium purge or vacuum
For light elements (e.g. aluminum or silicon), the ambient air absorbs a large proportion of the signal radiation. A helium purge or evacuation of the measurement chamber drastically improves the signal-to-noise ratio, noticeably lowers the detection limits compared with measurements in air, and thus optimizes measurement precision.
Practical test: How to select the right component for your requirements
Every measurement task has its own specific requirements for measurement technology. Which combination of detector, optics and measurement environment is right for you? The following table provides quick guidance for practical applications:
| Your requirement | Our recommendation |
|---|---|
Simple coating thickness measurements | A proportional counter tube can be a suitable detector option. |
More demanding coating thickness measurements | A Si-PIN detector offers higher energy resolution. |
Coating thickness measurements in the nanometer range or complex multilayer measurements | An SDD can provide the required performance. |
Material and elemental analysis | An SDD is particularly suitable for these applications. |
Small measurement spots | A microfocus X-ray tube and suitable X-ray optics / small collimators can be advantageous. |
Very small measurement spots | Polycapillary optics can focus measurement spots down to approximately 10 µm. |
Analysis of light elements | A helium purge or vacuum can improve the signal-to-noise ratio. |
Simple coating thickness measurements |
More demanding coating thickness measurements |
Coating thickness measurements in the nanometer range or complex multilayer measurements |
Material and elemental analysis |
Small measurement spots |
Very small measurement spots |
Analysis of light elements |
A proportional counter tube can be a suitable detector option. |
A Si-PIN detector offers higher energy resolution. |
An SDD can provide the required performance. |
An SDD is particularly suitable for these applications. |
A microfocus X-ray tube and suitable X-ray optics / small collimators can be advantageous. |
Polycapillary optics can focus measurement spots down to approximately 10 µm. |
A helium purge or vacuum can improve the signal-to-noise ratio. |
The correct configuration of your measuring instrument should always be selected based on the sample, measurement geometry, required measurement spot size, coating system, measurement mode and required detection limit.
FAQ
To optimally tailor your future XRF instrument to your measurement task, we recommend clarifying the following questions together with our FISCHER experts:
- What type of sample is to be measured?
Consider the material, size, geometry and surface condition of the sample. - What is to be measured?
Determine whether a coating thickness measurement, elemental analysis, or both are required. - What coating thickness range is required?
The required coating thickness range influences the selection of the detector and the configuration of the XRF instrument. - How large does the measurement spot need to be?
Small or complex measurement areas may require a microfocus X-ray tube, collimators or polycapillary optics. - Which elements must be detected?
The element range and required detection limit influence the excitation conditions and measurement environment. - Is a specific measurement direction required?
Measurement from top to bottom, bottom to top, or flexible measurement directions are suitable for different sample geometries. - Is air sufficient as the measurement environment?
For light elements, a helium purge or vacuum can improve measurement performance.
Conclusion
Consulting protects against costly purchasing mistakes
A high-performance XRF measuring instrument is an investment in the quality assurance of your company. But the “best” instrument is not categorically the most expensive model – rather, it is the one whose components and configuration are optimally suited to your individual measurement tasks. While an over-dimensioned instrument causes unnecessary costs, an under-dimensioned instrument leads to imprecise results. A carefully configured XRF instrument can therefore provide the required measurement results without causing unnecessary equipment and costs.
As a specialist in measurement technology with more than 70 years of experience, HELMUT FISCHER is at your side as a competent and fair partner, from the initial sample examination through to the purchase of your first instrument. We analyze your samples in advance in our application laboratories and recommend exactly the configuration you really need for your quality assurance. To ensure that you can enjoy your instrument for as long as possible and always obtain reliable results, we place particular emphasis on high-quality components and personal customer service with fast response times.